eBioscience ELISPOT Ready-SET-Go! reagent sets contain the necessary reagents for performing enzyme linked immunosorbent spot (ELISPOT) assays for high resolution frequency analysis of cytokine-secreting cells. These ELISPOT reagent sets are pre-titrated for optimal spot development and convenience.
eBioscience ELISPOT reagent sets are guaranteed to perform as specified for at least 12 months from date of receipt if stored and handled as specified by the accompanying datasheet and Certificate of Analysis.
For research use only, not for diagnostic or therapeutic use.
Capture Antibody: Pre-titrated, Functional Grade (low endotoxin) purified antibody
Detection Antibody: Pre-titrated, biotin-conjugated antibody
ELISA/ELISPOT Coating Buffer Powder (Reconstitute to 1L with dH20 and 0.22 uM filter)
Assay Diluent: 5X concentrated
Detection enzyme: pre-titrated Avidin-HRP
Certificate of Analysis: Lot-specific instructions for dilution of antibodies and enzyme
96 Well PVDF Membrane ELISPOT Plates (Millipore, Cat. No. MAIPS4510)
AEC substrate (Sigma, Cat. No. A-5754)
Distilled water
Wash Buffer: 1 x PBS, 0.05% Tween-20 (or eBioscience ELISA Wash Buffer Powder, cat 00-0400)
Pipettes and pipettors
Refrigerator
Incubator
Laminar Flow Hood
Plate Washer: Wash bottle or automated wash machine
ELISPOT plate reader or dissecting microscope for visual inspection
1 overnight antibody incubation
1-2 day cell activation
3-5 hr washing, antibody incubation, color development
Aseptic Procedures: (Use sterile buffers and aseptic conditions; use laminar flow hood for procedures.)
Dilute Functional Grade purified capture antibody in sterile ELISPOT Coating Buffer, as noted on Certificate of Analysis which is included with the reagent set. Coat ELISPOT plate with 100μl/well of capture antibody solution. Incubate at 4°C overnight.
Decant or aspirate coating antibody from plate.
Wash plates 2 times with 200μl/well sterile ELISPOT Coating Buffer. Decant.
Block plate with 200μl/well of complete RPMI-1640 at room temperature for 1 hour. Decant or aspirate plate.
Aliquot mitogen, antigen or controls diluted in complete RPMI-1640 medium to appropriate wells at 100μl/well. Aliquot cells at desired densities (e.g., 1x105/ml - 2x106/ml) at 100μl/well and incubate at 37°C, 5% CO2 humidified incubator for 24-48 hours.
Note: Kinetics and cell densities vary with target cytokine, treatment, and cell type and must be empirically determined. See references. Cells can be diluted in a sterile tissue culture plate starting at 2x106/well in triplicate wells with a series of 1:3 or 1:4 serial dilutions down the plate, and then transferred to the ELISPOT plate.
Non-Aseptic Procedures
Decant cells and medium from plates. Wash plate 3 times with ELISPOT Wash Buffer.
Dilute biotinylated detection antibody in Assay Diluent according to instructions on the Certificate of Analysis provided with the reagent set. Add 100μl/well to plate microwells and incubate at room temperature for 2 hrs (or 4°C overnight).
Decant antibody solution. Wash 4 times with ELISPOT Wash Buffer. Allow wells to soak for 1 minute for each wash.
Dilute Avidin-horseradish peroxidase reagent in Assay Diluent according to instructions on the Certificate of Analysis provided with the reagent set. Add 100μl/well of AV-HRP and incubate at room temperature for 45 minutes.
Decant Av-HRP solution. Wash plate 3 times with ELISPOT Wash Buffer, and then 2 times with 1x PBS (no Tween-20).
Add 100μl/well of freshly-prepared AEC Substrate Solution and develop at room temperature for 10-60 minutes; monitor development of spots.
Stop the substrate reaction by washing wells 3 times with 200μl/well distilled water.
Air-dry the plate. Count spots using a dissecting microscope or automated ELISPOT plate reader. Store plates in the dark prior to reading.
ELISA/ELISPOT Coating Buffer Powder
Reconstitute powder to 1L in dH20; sterile filter using 0.22 uM filter
Complete RPMI-1640:
RPMI-1640 with 10% FBS and 1% Pen/Strep/L-Glu
Assay Diluent (supplied as 5X):
Dilute 5X solution to 1X in DI H2O
ELISPOT Wash Buffer:
1X PBS with 0.05% Tween-20 (0.5 ml Tween-20 in 1 L PBS) or eBioscience ELISA Wash Buffer Powder
1X PBS:
80.0g NaCl
11.6g Na2HPO2
2.0g KH2PO4
2.0g KCl
Qs with DI H20 up to 10.0L, pH to 7.0
AEC (3-amino-9-ethyl carbazole) Substrate Solution:
AEC Stock Solution: Dissolve 100mg of AEC in 10ml of N,N Dimethylformamide (DMF; Pierce, Cat. No. 20672)
Add 333μl of AEC Stock Solution to 10ml of 0.1 M Acetate Solution (pH 5.0) (see below for recipe). Filter through a 0.45μm filter.
Just before use, add 5μl of 30% H2O2. Mix and use immediately.
0.1 M Acetate Solution (pH 5.0):
Combine 148ml 0.2 M acetic acid (11.55 ml glacial acetic acid per 1L dH2O) with 352ml 0.2 M sodium acetate (27.2 g sodium acetate per 1L dH2O).
Qs to 1L with dH2O. Adjust pH to 5.0.
Gebauer BS, et al. 2002. Evolution of the enzyme-linked immunosorbent spot assay for post-transplant alloreactivity as a potentially useful immune monitoring tool. Am. J. Transplant. 9: 857-866.
Guerkov RE, et al. 2003. Detection of low-frequency antigen-specifi c IL-10-producing CD4(+) T cells via ELISPOT in PBMC: cognate vs. nonspecifi c production of the cytokine. J. Immunol. Methods. 279: 111-121.
Kreher CR, et al. 2003. CD4+ and CD8+ cells in cryopreserved human PBMC maintain full functionality in cytokine ELISPOT assays. J. Immunol. Methods. 278: 79-93.
Ott PA, et al. 2004. CD28 costimulation enhances the sensitivity of the ELISPOT assay for detection of antigen-specifi c memory effector CD4 and CD8 cell populations in human diseases. J. Immunol. Methods. 285: 223-235.
Smith JG, et al. 2001. Development and validation of a gamma interferon ELISPOT assay for quantitation of cellular immune responses to varicella-zoster virus. Clin. Diag. Lab. Immunol. 8: 871-879.
Shafer-Weaver K, et al. 2003. The Granzyme B ELISPOT assay: an alternative to the 51Cr-release assay for monitoring cell-mediated cytotoxicity. J. Translational. Med. 1: 14.
Rininsland F, et al. 2000. Granzyme B ELISPOT assay for ex vivo measurements of T cell immunity. J. Immunol. Meth. 240:143-155.